Sentry Industries 5000 User Manual

Page of 130
SMC
 sierra monitor corporation 
Sentry Instruction Manual - Version 6 
 
SERVICE
 
Page:   70 
3.  Using a Sierra Monitor Model 1200-26 Gas 
Calibrator with a Sierra Monitor Model 5358-00 
calibration adapter, or Model 5360 Gas 
Delivery Fitting. Apply the calibration gas at a 
flow rate of 100 cc/min. until the signal out 
voltage stops changing (approx. 2 minutes).  
Then adjust span adj. potentiometer to the 
calculated value. 
4.  Remove the calibration gas and wait for the 
voltage to drop.  If the voltage is below 0.30 
VDC the procedure is complete.  If the voltage 
is above 0.30 VDC repeat the zero and span 
adjustments. 
5.  Calibrate the sensor using the instruction manual 
procedure. 
• 
LOW SENSITIVITY
 message during 
calibration indicates one of the two 
conditions for a combustible sensor: 
• 
The zero gas voltage must be less than 
.3 volts.  
• 
The new span change (span voltage less 
zero voltage) must exceed 30% of the 
calculated span change. 
For Example: 
Zero gas voltage = .20 VDC 
Span voltage = .43 VDC 
Span change is .43-.2  = .23 VDC 
As the calculated span value (from step 
5 above) is 1.03, the calculated span 
change is:  1.03-.2 = .87 
By calculation: 30% of .87 = .26 
The new span change (.23) is less than 
30% of the calculated span change so a 
LOW SENSITIVITY
 message will be 
displayed. 
The 
LOW SENSITIVITY
 calculation is made to insure that 
the calibration does not adjust the electronics so that the 
"gain" is large enough to cause false alarms due to minor 
drift or due to minimal electronic noise. 
7.5.4  SENSOR REPLACEMENT 
If any of the above adjustments cannot be completed 
successfully the sensor and/or electronics will require 
replacement. To check the sensor unplug it from the 
electronics and remove from the housing so that continuity 
and resistance tests can be made.  Table 7-4 provides the 
pin out for the sensor assembly. 
PIN 
FUNCTION 
WIRE COLOR 
DETECTOR 
WHITE 
COMPENSATOR 
BROWN 
CALIB. SW. 
RED 
DETECTOR 
BLACK 
DRAIN 
BARE 
CALIB. SW. 
GREEN 
Table 7-5 
Combustible Gas Sensor 
Wiring Pin Out 
1.  Detector element  resistance should be between 1.5 to 
2 ohms. 
2.  Compensator element resistance should be between 
1.5 to 2 ohms.  
3.  Calibration switch should be open in normal 
operation and should close when the calibration 
magnet is applied. 
If the sensor fails any of these tests it should be replaced.  
If the sensor passes these tests the electronics are suspect 
and BOTH the sensor and electronics should be replaced.  
(The electronics and sensor which are removed should be 
returned to the factory for evaluation). 
When a new sensor is installed the following actions must 
be taken: 
• 
Make the bridge voltage adjustment 
immediately to avoid over-voltage damage to 
the sensor. 
• 
Use diagnostic code 0020 to establish nominal 
value calibration. 
• 
Allow 24 hours for full stabilization of the 
sensor, recheck the zero and span adjustments 
and calibrate the sensor module.