Sentry Industries 5000 User Manual

Page of 130
SMC
 sierra monitor corporation 
Sentry Instruction Manual - Version 6 
 
SERVICE
 
Page:   81 
7.10  AMMONIA SENSOR MODULE (5100-25) 
7.10.1  DESCRIPTION 
The Ammonia Sensor Module includes a sensor 
assembly and electronic assembly installed in an 
explosion proof housing. 
The sensor assembly includes a rechargeable 
electrochemical sensor.  The assembly screws into 
one hub of the sensor module enclosure and plugs 
into the bottom electronics card via a six pin 
connector.   
Cabling from the controller connects to a three pin 
spring loaded terminal strip on the electronics 
assembly. 
7.10.2  TROUBLE ANALYSIS 
Electrical adjustment, recharge, or replacement of 
the sensor will be necessary under the following 
conditions: 
• 
Controller displays the following error 
messages 
SENSOR FAILURE 
LOW SENSITIVITY 
• 
False readings or alarms are received 
due to sensor inaccuracy. 
Warning:  During sensor adjustments the 
concentration reading on the controller will be 
inaccurate and alarm level concentrations may be 
displayed.  If false activation of the alarm relays will 
cause a problem disconnect the relay wiring prior to 
adjustment or turn the module off using the  "Change 
Module" mode. 
NOTE 
Although all the necessary data can be collected with a 
voltmeter at the sensor module, some helpful 
information can be displayed on the controller or 
printed.  See diagnostic codes 0004 and 0008 in 
Appendix C. 
7.10.3  ADJUSTMENT PROCEDURE 
Prior to reading voltages and making adjustments perform 
a visual inspection to confirm that there are no physical 
problems such as water  in the electronics enclosure, 
wiring damage or corrosion. 
Use Figure 7-13 to locate test points during the following 
procedures. 
7.10.3.1  SIGNAL ADJUSTMENT 
It is unlikely that any electrical adjustment will be 
required except when a new sensor is installed. 
To make the electrical adjustment connect a voltmeter to 
SIGNAL OUT
 (pos) and 
GND
 (ground) and use 
SPAN ADJ
 
potentiometer to adjust the voltage based on the 
following: 
• 
1% of full scale  = 0.02 VDC. 
• 
100% of full scale = 2.00 VDC. 
7.10.3.2  SENSOR RECHARGE 
The Ammonia sensor can be recharged by removing it 
from service, draining electrolyte, cleaning electrodes and 
replacing the electrolyte and membrane. 
Step by step instructions for sensor recharge are supplied 
with the recharge kit. 
7.10.3.3  SENSOR REPLACEMENT 
The Ammonia sensor assembly should be replaced when it 
can no longer be recharged and calibrated correctly. 
 
Figure 7-13 
Cover Plate - Ammonia Module