SERVICE
 
Page:   82 
The 
SENSOR FAILURE
 message for ammonia sensors does 
not indicate a specific failure of the sensor but indicates 
that the sensor is not correctly connected to the 
electronics.  If this message appears check that the sensor 
harness is correctly installed to the connector on the 
bottom electronics board. 
To replace the sensor assembly: 
1.  Confirm that system power has been removed. 
2.  Remove the transmitter electronics board from the 
main housing and unplug the sensor harness from 
the transmitter electronics. 
3.  Unscrew the sensor housing from the bottom of 
the enclosure 
4.  Install the new sensor assembly into the enclosure 
and tighten firmly. 
5.  Reconnect the sensor harness to the transmitter, 
install the transmitter into the housing and restore 
power. 
6.  Allow one hour for the sensor to stabilize prior to 
recalibration. 
7.  Recalibrate the sensor module.