C A L I B R A T I O N   A N D   T R O U B L E S H O O T I N G
Error Messages
04_06
LDX-3232
 101
 
C H A P T E R  
6
Error Messages
In the event of a hardware error condition, error messages will be displayed on the 
display. In most cases, the error message will appear for three seconds. In some 
cases the error code display will remain until the user changes display modes. In 
the case of multiple error messages, the display may sequentially show each 
message for three seconds. In addition to the hardware errors, GPIB errors may 
be read via the ERR? query. Table 3.4 lists the numerical error ranges by function. 
Table 3.5 contains all of the error messages that may be generated. Not all of 
these messages may appear on the front panel displays. Some refer to GPIB 
activities only.
In remote operation, the errors can be read by issuing the ERR? query. When this 
is done, all of the error messages that are resident in the error queue are returned 
(up to 10 may be stored). Reading the error queue via GPIB clears the error 
queue.
Table 6.3  Error Code Classifications
Table 6.4  LDX-3232 Series Error Codes
Error Code Range
Area of Operation
E-001 to E-099
Internal Program Errors
E-100 to E-199
Parser Errors
E-200 to E-299
Execution Control Errors
E-300 to E-399
GPIB Errors
E-500 to E-599
LASER Control Errors
Error Code
Explanation
E-001
Memory allocation failure
E-103
<DEFINITE LENGTH ARBITRARY BLOCK PROGRAM DATA> length too long
E-104
<NON-DECIMAL NUMERIC PROGRAM DATA> type not defined
E-105
<DECIMAL PROGRAM DATA> exponent not valid
E-106
<DECIMAL PROGRAM DATA> digit expected
E-123
Command not found
E-124
<program mnemonic> Lookup, failed because query/command type match failed
E-126
Too few or too many program data elements
E-201
<PROGRAM DATA> value out of range
E-202