C A L I B R A T I O N   A N D   T R O U B L E S H O O T I N G
Error Messages
102 
LDX-3232
C H A P T E R  
6
 
Error Code
Explanation
E-203
Security violation, command is not available without clearance
E-205
<PROGRAM DATA> is not a Boolean value or word
E-207
<PROGRAM DATA> will not convert to an unsigned 16-bit value
E-209
<PROGRAM DATA> will not convert to an unsigned 32-bit value
E-210
<PROGRAM DATA> will not convert to a floating point value
E-211
<PROGRAM DATA> will not convert to a character value
E-213
<PROGRAM DATA> is incorrect block data length
E-214
<PROGRAM DATA> length exceeds maximum
E-302
Query error; device was addressed to talk, but GPIB controller failed to read all of the 
<RESPONSE MESSAGE>
E-501
LASER interlock disabled output
E-503
LASER open circuit disabled output
E-504
LASER current limit disabled output
E-505
LASER voltage limit disabled output
E-507
LASER power limit disabled output
E-510
LASER out of tolerance disabled output
E-511
LASER control error disabled output
E-512
Analog section status is all 1’s or all 0’s (power down)
E-513
Serial EEPROM checksum error
E-515
Laser output must be off to change ranges
E-516
Incorrect configuration for calibration sequence to start
E-519
Setting a measurement is only valid during the calibration phase for that measurement; 
user has tried to calibrate a measurement without first entering the required calibration 
mode.