Accu-Chek insulin pump User Manual
Start-up graph over the stabilization period
The start-up graph shows changes in the flow rate over the stabilization time.
The measurements were made at a medium basal rate of 1.0 U/h according to
IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room
temperature.
IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room
temperature.
Influence of height on delivery accuracy
The average deviation of the delivered amount (overall mean percentage flow
error) for U100 insulin is
The average deviation of the delivered amount (overall mean percentage flow
error) for U100 insulin is
≤
± 10% when your ACCU-CHEK Spirit insulin pump is
located 3.3 feet (1 m) above/below the level of infusion site.
Appendixes
ACCU-CHEK Spirit Insulin Pump
174