Start-up graph over the stabilization period
The start-up graph shows changes in the flow rate over the stabilization time.
The measurements were made at a medium basal rate of 0.1 U/h according to
IEC 60601-2-24:1998 with a Disetronic Classic infusion set PC 16/110 at room
temperature.
Bolus
For U100 insulin the delivery accuracy of a maximum bolus is 
± 5% and the
delivery accuracy of a minimum bolus is 
± 30%.
The measurements were made according to IEC 60601-2-24:1998 with a
Disetronic Classic infusion set PC 16/110 at room temperature.
Delivery Mode
Delivery Rate/Volume
Accuracy
Range
Basal
0.1 U/h
±20%
0.08 - 0.12 U/h
Bolus
0.1 U
±30% 
0.07 - 0.13 U
Appendixes
ACCU-CHEK Spirit Insulin Pump