Accu-Chek insulin pump User Manual

Page of 200
Typical Range and
Maximum Parameter 
Parameter Settings
Range
Basal Rate
– hourly basal rate + 
U100: 62.5 IU/h
Insulin Type
Max. Amount
max. TBR
U100
0 – 62.50 IU/h
(hourly basal rate 
combined with temporary 
basal rate, max. increase)
– hourly basal rates 
0.1 U
increment
– hourly basal rates 
U100: 10.0 IU/h
Insulin Type
Max. Amount
U100
0 – 25.0 IU/h
– temporary basal rate
100%
0 – 200%
– TBR increase or decrease
10%
10%
in increments of
– TBR duration
max. 24 h
15 min – 24 h
– TBR duration increments
15 min
15 min, 30 min, 1 h
– basal rate profiles available
5
1 – 5
– basal rate lock
off
on or off
Bolus
– bolus amount per
U100: 25.0 IU/h
Insulin Type
Max. Amount
bolus delivery
U100
0 – 25.0 U
– bolus amount increments
0.5U
0.1, 0.2, 0.5, 1.0 or 2.0 U
for “Quick” and “Scroll”
Standard Bolus
– bolus duration (Extended
max. 24 h
15 min – 24 h
bolus, MultiWave bolus)
– bolus duration increments
15 min
15 min, 30 min, 1h
Appendixes
ACCU-CHEK Spirit Insulin Pump
178