DOD PS-7 User Manual

Page of 9
 
 
H – H mode (1
st
 level: Upper limit, 2
nd
 level: Upper limit) 
 
 
 
 
 
                                                   
  Normal state (No gas is detected) 
 
 
L – L mode (1
st
 level: Lower limit, 2
nd
 level: Lower limit) 
 
 
                                                      Normal state   
 
 
 
 
 
            H – L mode (Upper limit, Lower limit) 
 
 
 
                                                                    Normal state 
 
 
 
 
 
* A2 value should be always greater than A1 value. 
 
 
 
 
 
 
 
 
 
 
25vol% 
21vol% 
1
st
 level alarm : ALARM 1 is displayed.
Alarm contact ZA1 is activated. 
Alarm value is set by entering A2 **
0 vol% 
2
nd
 level alarm : ALARM 1 , ALARM 2 is displayed.
Alarm contact ZA1 and ZA2 are activated. 
Alarm value is set by entering A1 **
 
0ppm 
Full Scale 
2
nd
 level gas alarm : ALARM 1 , ALARM 2 is displayed.
Alarm contacts ZA1 and ZA2 are activated 
Alarm value is set by entering A2 ** 
 
1
st
 level gas alarm : ALARM 1    is displayed. 
Alarm contact ZA1 is activated 
Alarm value is set by entering A1 ** 
Upper alarm : ALARM 2 is displayed.
Alarm contact ZA2 is activated. 
Alarm value is set by entering A2 **
50vol% 
21vol% 
0 vol% 
Lower alarm : ALARM 1 is displayed.
Alarm contact ZA1 is activated. 
Alarm value is set by entering A1 **