Multi-Echo EPI 
Page 4 of 9 
Numaris/4 VA21B 
1.2  Image contrast of Multi-Echo EPI 
 
The signal intensity of the single echo images of the Multi-Echo EPI sequence is 
simply given by: 
 
*)
2
/
exp(
0
T
TE
S
S
=
 
 
where  S0 is the signal intensity at echo time 0, which corresponds to the proton 
density for long repetition times TR, and TE is the echo time. Therefore, the echo 
times of the different echo-images should be on the order or less than the tissue T2*. 
Parametric images of the proton density and T2* can be obtained by fitting a model 
function to the signal intensities at the different echo times (3). Please see the 
processing options described below. 
 
 
 
 
Figure 2 – Head images acquired on Magnetom TRIO with the Multi-Echo EPI 
sequence with 4 different echo times in a single shot. 
 
 
1.3 Processing options 
 
Several processing options are given in the method. The different echo images are 
always saved separately as the original images and optionally parametric maps can 
be calculated in the reconstruction. In addition, online 3D-motion correction (even 
prospective – PACE) and t-test can be performed. Please see sequence and 
protocols for details. 
 
 
1.4 References 
 
1.  P. Mansfield, A.A. Maudsley. Planar Spin Imaging. J. Magn. Reson. 27, 101 
(1977) 
2.  O. Speck, J. Hennig. Functional Imaging by I
0
- and T
2
*-Parameter Mapping 
using Multi-Image-EPI. Magn. Reson. Med. 40(2), 243-248 (1998)