DR4300 Circular Chart Recorder Product Manual
12/03
Procedure
The procedure for calibrating the current proportional output is listed in Table 5-11. Make sure “LOCK” in
the Set Up group is set to “NONE.”
the Set Up group is set to “NONE.”
For display recorders with 2 pens, be sure the correct input is on display—INP 1 or INP 2 indicator is lit.
Press
FUNC
key to change input on display to agree with the control loop output to be calibrated.
Table 5-11 Procedure for Calibrating Current Output
Step Description
Press
Action/Result
1
Enter Calibration Mode
SET
UP
UP
until you see
CAL
Upper Display
CURENT
Lower Display
If “CAL” doesn’t appear or prompt access is denied, check
the “LOCK” configuration.
the “LOCK” configuration.
2
Calibrate 0 %
FUNC
You will see:
Upper Display
a value (approximately 365)
ZERO
Lower Display
or
Until the desired 0 % output is read on the milliammeter.
Use the values shown below depending on the action of
your recorder.
Use the values shown below depending on the action of
your recorder.
3
Calibrate 100 %
FUNC
Stores the 0 % value and, you will see:
Upper Display
a value (approximately 1800)
SPAN
Lower Display
or
Until the desired 100 % output is read on the milliammeter.
Use the values shown below depending on the action of
your recorder.
Use the values shown below depending on the action of
your recorder.