Agilent Technologies E7405A User Manual

Page of 92
Chapter 1
37
Making Basic Measurements
Measuring Low Level Signals
9.
Place the signal at center frequency by pressing 
Peak Search
Marker
Mkr
CF
10.
Press 
BW/Avg
Res BW
, and then 
. The low level signal appears 
more clearly because the noise level is reduced. As shown in 
#
 mark appears next to the 
Res BW
 annotation at the lower left 
corner of the screen, indicating that the resolution bandwidth is 
uncoupled. As the resolution bandwidth is reduced, the sweep time 
is increased to maintain calibrated data. 
Figure 1-29
Decreasing Resolution Bandwidth 
Measuring Low Level Signals Example 3:
Narrowing the video filter can be useful for noise measurements and 
observation of low level signals close to the noise floor. The video filter 
is a post-detection low-pass filter that smooths the displayed trace. 
When signal responses near the noise level of the analyzer are visually 
masked by the noise, the video filter can be narrowed to smooth this 
noise and improve the visibility of the signal. (Reducing video 
bandwidths requires slower sweep times to keep the analyzer 
calibrated.) 
Using the video bandwidth function, measure the amplitude of a low 
level signal.
1.
Connect a signal generator to the analyzer input. 
2.
Set the signal generator frequency to 300 MHz with an amplitude of 
80 dBm.