Agilent Technologies E7405A User Manual

Page of 92
Chapter 1
39
Making Basic Measurements
Measuring Low Level Signals
NOTE
The video bandwidth must be set wider than the resolution bandwidth 
when measuring impulse noise levels.
Figure 1-31
Decreasing Video Bandwidth 
Measuring Low Level Signals Example 4:
If a signal level is very close to the noise floor, video averaging is 
another way to make the signal more visible. 
NOTE
The time required to construct a full trace that is averaged to the 
desired degree is approximately the same when using either the video 
bandwidth or the video averaging technique. The video bandwidth 
technique completes the averaging as a slow sweep is taken, whereas 
the video averaging technique takes many sweeps to complete the 
average. Characteristics of the signal being measured, such as drift and 
duty cycle, determine which technique is appropriate. 
Video averaging is a digital process in which each trace point is 
averaged with the previous trace-point average. Selecting 
Average Type
Video Avg
 and 
Average (On)
 changes the detection mode from peak to 
sample. The result is a sudden drop in the displayed noise level. The 
sample mode displays the instantaneous value of the signal at the end 
of the time or frequency interval represented by each display point, 
rather than the value of the peak during the interval. Sample mode is 
not used to measure signal amplitudes accurately because it may not 
find the true peak of the signal. 
Video averaging clarifies low-level signals in wide bandwidths by 
averaging the signal and the noise. As the analyzer takes sweeps, you 
can watch video averaging smooth the trace.