Chapter 1
49
Making Basic Measurements
Measuring Signal-to-Noise
Measuring Signal-to-Noise 
The signal-to-noise measurement procedure below may be adapted to 
measure any signal in a system if the signal (carrier) is a discrete tone. 
If the signal in your system is modulated, it will be necessary to modify 
the procedure to correctly measure the modulated signal level
In this example the 50 MHz amplitude reference signal is used as the 
fundamental source. The amplitude reference signal is assumed to be 
the signal of interest and the internal noise of the analyzer is measured 
as the system noise. To do this, you will need to set the input attenuator 
such that both the signal and the noise are well within the calibrated 
region of the display. 
Signal-to-Noise Measurement Example:
Perform the steps below to measure the signal-to-noise.
1.
Perform a factory preset by pressing 
Preset
Factory Preset
 (if 
present).
2.
Turn on the internal 50 MHz amplitude reference signal of the 
analyzer as follows: 
For the E7401A, use the internal 50 MHz amplitude reference 
signal of the analyzer as the signal being measured. Press 
Input/Output
Amptd Ref (On)
For all other models connect a cable between the front-panel 
AMPTD REF OUT to the analyzer INPUT, then press 
Input/Output
Amptd Ref Out
 
(On)
.
3.
Set the center frequency to 50 MHz by pressing 
FREQUENCY
Center Freq
50
MHz
4.
Set the span to 1 MHz by pressing 
SPAN
Span
, 1, 
MHz
.
5.
Set the Y-Axis Units to dBm by pressing 
AMPLITUDE
More, 
Y-Axis Units,
 
dBm
6.
Set the resolution bandwidth to spectrum analyzer coupling by 
pressing 
BW/Avg
Res BW
 
(SA)
7.
Set the reference level to –10 dBm by pressing 
AMPLITUDE
Ref Level
–10, 
dBm
.
8.
Set the attenuation to 40 dB by pressing 
AMPLITUDE
Attenuation
, 40, 
dB
.
9.
Press 
Peak Search
 to place a marker on the peak of the signal.
10.
Press 
Marker
Delta
, 200, 
kHz
 to put the delta marker in the noise at 
the specified offset, in this case 200 kHz.