Chapter 1
51
Making Basic Measurements
Making Noise Measurements
Making Noise Measurements
There are a variety of ways to measure noise power. The first decision 
you must make is whether you want to measure noise power at a 
specific frequency or the total power over a specified frequency range, 
for example over a channel bandwidth. 
Noise Measurement Example 1:
Using the marker function, 
Marker Noise
, is a simple method to make a 
measurement at a single frequency. In this example, attention must be 
made to the potential errors due to discrete signal (spectral 
components). This measurement will be made near the 50 MHz 
amplitude reference signal to illustrate the use of 
Marker Noise.
1.
Perform a factory preset by pressing 
Preset
Factory Preset
 (if 
present).
2.
Turn on the internal 50 MHz amplitude reference signal of the 
analyzer as follows: 
For the E7401A, use the internal 50 MHz amplitude reference 
signal of the analyzer as the signal being measured. Press 
Input/Output
Amptd Ref
 (On). 
For all other models connect a cable between the front-panel 
AMPTD REF OUT to the analyzer INPUT, then press 
Input/Output
Amptd Ref Out
 (On).
3.
Set the center frequency to 49.98 MHz by pressing 
FREQUENCY
Center Freq
, 49.98, 
MHz
4.
Set the span to 100 kHz by pressing 
SPAN
Span
, 100, 
kHz
.
5.
Set the resolution bandwidth to 1 kHz by pressing 
BW/Avg
Res BW
(Man), 1, 
kHz
.
6.
Set the Y-Axis Units to dBm by pressing 
AMPLITUDE
More, 
Y-Axis Units,
 
dBm
7.
Set the attenuation to 60 dB by pressing 
AMPLITUDE
Attenuation
 
(Man)
, 60, 
dB
. See 
NOTE
When making noise measurements and 
AMPLITUDE
Scale Type
 (Log) is 
selected (10 dB/division), position the trace between 3 and 6 graticule 
lines above the bottom by adjusting the reference level (
AMPLITUDE
Ref Level
). Measurement inaccuracies may occur if displayed trace is 
positioned outside this range.