Agilent Technologies E7405A User Manual

Page of 92
76
Chapter 2
 
Making Complex Measurements
Making Stimulus Response Measurements
Tracking Generator Unleveled Condition 
When using the tracking generator, the message 
TG unleveled
 may 
appear. The 
TG unleveled
 message indicates that the tracking 
generator source power (
Source
Amplitude
) could not be maintained at 
the selected level during some portion of the sweep. If the unleveled 
condition exists at the beginning of the sweep, the message will be 
displayed immediately. If the unleveled condition occurs after the sweep 
begins, the message will be displayed after the sweep is completed. A 
momentary unleveled condition may not be detected when the sweep 
time is short. The message will be cleared after a sweep is completed 
with no unleveled conditions. 
The unleveled condition may be caused by any of the following: 
• Start frequency is too low or the stop frequency is too high. The 
unleveled condition is likely to occur if the true frequency range 
exceeds the tracking generator frequency specification (especially 
the low frequency specification). 
• Source attenuation may be set incorrectly (select 
Attenuation
 (Auto) 
for optimum setting). 
• The source power may be set too high or too low, use 
Amplitude
 (Off) 
then 
Amplitude
 (On) to reset it. 
• The source power sweep may be set too high, resulting in an 
unleveled condition at the end of the sweep. Use 
Power Sweep
 (Off) 
then 
Power Sweep
 (On) to decrease the amplitude. 
• Reverse RF power from the device under test detected by the 
tracking generator ALC (automatic level control) system.
Measuring Device Bandwidth 
It is often necessary to measure device bandwidth, such as when testing 
a bandpass filter. There is a key in the 
Peak Search
 menu that will 
perform this function. The device signal being measured must be 
displayed before activating the measurement. The span must include 
the full response.
Activate the measurement by toggling the 
N dB Points
 key to On. The 
analyzer places arrow markers at the 
3 dB points on either side of the 
response and reads the bandwidth. For other bandwidth responses 
enter the number of dB down desired, from 
1 dB to 
80 dB.
No other signal can appear on the display within N dB of the highest 
signal. The measured signal cannot have more than one peak that is 
greater than or equal to N dB. A signal must have a peak greater than 
the currently defined peak excursion to be identified. The default value 
for the peak excursion is 6 dB.