Agilent Technologies E7405A User Manual

Page of 92
Chapter 2
77
Making Complex Measurements
Making Stimulus Response Measurements
Measurements are made continuously, updating at the end of each 
sweep. This allows you to make adjustments and see changes as they 
happen. The single sweep mode can also be used, providing time to 
study or record the data.
The N dB bandwidth measurement error is typically 
±
1% of the span.
Example:
Measure the 3 dB bandwidth of a 200 MHz bandpass filter.
1. To measure the rejection of a bandpass filter, connect the equipment 
as shown in 
. This example uses a 200 MHz bandpass 
filter. 
Figure 2-5 
Transmission Measurement Test Setup 
2. Perform a factory preset by pressing 
Preset
Factory Preset
 (if 
present). 
3. Set the Y-Axis Units to dBm by pressing 
AMPLITUDE
More, 
Y-Axis Units,
 
dBm
4. Set the span to 100 MHz by pressing 
SPAN,
 
Span
, 100, 
MHz
.
5. Set the center frequency to 200 MHz by pressing 
FREQUENCY,
 
Center Freq
, 200, 
MHz
.
6. Set the resolution bandwidth to 10 kHz by pressing 
BW/Avg
Res BW
10, 
kHz
7. Turn on the tracking generator and if necessary, set the output 
power to –10 dBm by pressing 
Source
Amplitude
 (On), –10, 
dBm
CAUTION
Excessive signal input may damage the DUT. Do not exceed the 
maximum power that the device under test can tolerate.