Chapter 2
 
Making Complex Measurements
Making a Reflection Calibration Measurement
Making a Reflection Calibration Measurement 
The calibration standard for reflection measurements is usually a short 
circuit connected at the reference plane (the point at which the device 
under test (DUT) will be connected.) Se
. A short circuit 
has a reflection coefficient of 1 (0 dB return loss). It reflects all incident 
power and provides a convenient 0 dB reference. 
Figure 2-13 
Reflection Measurement Short Calibration Test Setup 
Example:
Measure the return loss of a filter. The following procedure makes a 
reflection measurement using a coupler or directional bridge. This 
example uses a 200 MHz bandpass filter as the DUT. 
NOTE
The analyzer center frequency and span for this measurement can 
easily be set up using the transmission measurement setup in 
. Tune the analyzer so 
that the passband of the filter comprises a majority of the display, then 
proceed with the steps outlined below.