Chapter 2
85
Making Complex Measurements
Making a Reflection Calibration Measurement
Reflection Calibration 
1. Connect the DUT to the directional bridge or coupler as shown in 
Terminate the unconnected port of the DUT. 
NOTE
If possible, use a coupler or bridge with the correct test port connector 
for both calibrating and measuring. Any adapter between the test port 
and DUT degrades coupler/bridge directivity and system source match. 
Ideally, you should use the same adapter for the calibration and the 
measurement. Be sure to terminate the second port of a two port device. 
2. Connect the tracking generator output of the analyzer to the 
directional bridge or coupler. 
3. Connect the analyzer input to the coupled port of the directional 
bridge or coupler. 
4. Perform a factory preset by pressing 
Preset
Factory Preset
 (if 
present). 
5. Set the Y-Axis Units to dBm by pressing 
AMPLITUDE
More, 
Y-Axis Units,
 
dBm
6. Turn on the tracking generator and if necessary, set the output 
power to –10 dBm by pressing 
Source
Amplitude
 (On), –10, 
dBm
CAUTION
Excessive signal input may damage the DUT. Do not exceed the 
maximum power that the device under test can tolerate. 
7. Set the span to 100 MHz by pressing 
SPAN,
 
Span
, 100, 
MHz
.
8. Set the center frequency to 200 MHz by pressing 
FREQUENCY,
 
Center Freq
, 200, 
MHz
.
9. Set the resolution bandwidth to 3 MHz by pressing 
BW/Avg
Res BW
3, 
MHz
10.Replace the DUT with a short circuit. 
11.Normalize the trace by pressing 
Trace/View
More
Normalize
Store Ref
 (1
3), 
Normalize
 (On). See 
.
This will activate the trace 1 minus trace 3 function and display the 
results in trace 1. The normalized trace or flat line represents 0 dB 
return loss. Normalization occurs each sweep. Replace the short 
circuit with the DUT.
NOTE
Since the reference trace is stored in trace 3, changing trace 3 to 
Clear Write will invalidate the normalization.