CFA­3000 ISE Analyzer
Instruction Manual
2.8   Prepare Reagent Containers
Prepare reagent containers as follows: 
1.   Remove screw caps from reagent bottles.
2.   Using an awl, make one hole in foil for the reagent straw. Do not allow awl to touch reagent or standards.
3.   Place Standard and Reagent bottles into reagent cabinet, as shown in Figure 2­8.
4.   Insert color­coded reagent straws and labeled Standard straw leading from valve pump assembly into holes    
punched in corresponding bottles. 
Figure 2­8. Sample Bottle Placement in Cabinet (for all methods; colors refer to reagent straw colors)