Emerson Process Management CFA-3000 User Manual

Page of 74
25
CFA3000 MANUAL
SECTION 3
INSTRUMENT DESCRIPTION
Place the clear grab sample inlet tube into a beaker containing the grab sample. 
2.   Press                one time. The display shows flashing dashes,  “ ­­ ­­ ­­ ­­“ .
The system analyzes the grab sample and after 15 minutes, displays the results when complete (The results
are “flashed” on the display).
Grab Sample Display:  10.0      (flashing for three minutes)
After the display of the grab sample’s value, a multi­stream analyzer begins processing the next stream in sequence. 
3.   To display measurement for previous grab sample, press               twice. 
4.   Press              button to return to normal operation.          
3.6   Manually Sample a Particular Stream. 
Use this procedure to manually switch between sample streams in a multi­stream system. This operation can be
performed at any time, except during calibration. 
Press   one time.                
Sample Display:  0 0 0 1 .
Press               and               to change stream number.     Sample Display:  0 0 0 3
Press              to start sampling desired stream.
Press              to return to multi­stream automatic analysis.
3.7   Shutdown 
To shut down the CFA­3000 for only a few days, press main Power button in back of upper cabinet to OFF
position. Leave reagent straws in reagent containers.