Emerson Process Management CFA-3000 User Manual

Page of 74
26
CFA3000 MANUAL
SECTION 3
INSTRUMENT DESCRIPTION
To shut down the analyzer for more than a day, perform the following procedure
1.  Remove straws from reagent containers and place them into a beaker of DI water.
2.  Press             twice within a two­second interval to activate priming. This allows wash solution or DI    
water to be injected rapidly through system. 
3.   After priming finishes (after 15 minutes or when “auto cal” starts), press the main Power button in back of  
upper cabinet to OFF position. 
3.8   Quick Start
If the analyzer has been shut down during normal operations, turn power on and 
press              once.
If the CFA­3000 has been turned off for an extended shutdown:
Press              twice to prime; the system will automatically go into :”Auto Cal” after priming
Some chemistries may require specific wash solutions. Check Appendix A, 
"Chemistries," for details.