dhi pk-7000-mpc1-d interconnections kit installation instructions Manual De Usuario
Calibration Solutions
for Pressure
and Flow™
PK-7000-MPC1-D
Interconnections Kit
Installation Instructions
INTERCONNECTIONS KIT, P/N 401645, INCLUDES
• 6 100319 Adaptor, 1/8 in. tube x 1/8 in. NPT M
• 2 102184
Reducer, 4P x 2P
• 2 100317 Adaptor, 1/8 in. tube x 1/4 in. NPT F
• 1 101988-Z Tubing, SS, .125 in. OD x 300 cm
• 2 401459 Quick Connect and Stand Assy
• 2 102116
Quick Connect, 1/8 in. NPT F
PARTS ON TEST PORT OR IN ACCESSORIES OF PG7000
• 1 101200 Collar
• 1 400518 Adaptor , 1/8 in. tube x DH 200 M
• 1 100272 Gland
TO PNEUMATICALLY CONNECT THE PG7000
S
OR RPM
S
TO THE MPC1-D
•
MPC1-D is intended for controlling pressures in a high line differential pressure system where there is a “high” side and
a “low”. The instructions below should be repeated, once for the high side and once for the low side. See the MPC1-D
Operation and Maintenance manual for a schematic of a typical system layout.
•
If you are using RPMs (not PG7000s) go directly to Step
p
b.
n
Remove gland, collar and plastic plug from the PG7000 TEST port.
o
Install gland and collar onto 1/8 in. tube x DH200 male adaptor supplied in PG7000 accessories.
Note that the collar threads left handed. Assure it is properly attached. The orange plastic plug is
not used.
Note that the collar threads left handed. Assure it is properly attached. The orange plastic plug is
not used.
pa PG7000 - Screw the adaptor (as assembled) into the PG7000 TEST port. Torque to 15 Nm (Newton
meter).
pb RPM3 - Screw 1/8 in. tube x 1/8 in. NPT male adaptor into the RPM3 TEST port using Teflon tape as
a seal.
TO MPC1-D
TEST HI or TEST LO
TEST HI or TEST LO
TO MPC1-D
REF HI or REF LO
REF HI or REF LO
101988-Z
102184
100319
100319
100317
102116
401459
101988-Z
400518
INCLUDED WITH P 7000
G
100272
100319
TO RPM3
101200
TO P 7000