Calibration 
 
The TorqueTrak 10K System is calibrated prior to shipping using 
instruments traceable to the United States National Institute of 
Standards and Technology (NIST). Calibration can be checked at 
any time with a NIST traceable reference such as a calibrated 
voltmeter with sufficient (millivolt) resolution. 
To verify calibration of the RX10K Receiver: 
1.  Insert connector on AC/DC adapter into Power Input  jack on 
the RX10K rear panel (refer to Figure 2 on page 6).  Plug 
adapter into appropriate AC power source (i.e., wall socket).  
Flip the RX10K power switch to On  while holding down the 
ADJUST 
▲ key.   
2.  Allow the RX10K to warm up for 15 minutes. 
3.  Connect a calibrated, high-accuracy voltmeter to the Voltage 
Output terminals. 
4.  Scroll RX10K display to Input parameter screen.  Press the 
ADJUST 
▲  key to scroll through the simulated inputs and 
check the outputs. 
Input 
Output  
+FS 
10.000 
± .010 VDC 
Zero 
0.000 
± .005 VDC 
-FS 
-10.000 
± .010 VDC 
+FS/2 
5.000 
± .005 VDC 
-FS/2 
-5.000 
± .005 VDC 
+FS/4 
2.500 
± .005 VDC 
-FS/4 
-2.500 
± .005 VDC 
 
It is recommended that the system be checked for calibration 
annually.  If found to be out of specification, it can be returned to 
Binsfeld Engineering Inc. for calibration for a nominal fee 
($150.00, price subject to change).