- 21 -
Calibration
The TorqueTrak 10K System is calibrated prior to shipping using
instruments traceable to the United States National Institute of
Standards and Technology (NIST). Calibration can be checked at
any time with a NIST traceable reference such as a calibrated
voltmeter with sufficient (millivolt) resolution.
To verify calibration of the RX10K Receiver:
1. Insert connector on AC/DC adapter into Power Input jack on
the RX10K rear panel (refer to Figure 2 on page 6). Plug
adapter into appropriate AC power source (i.e., wall socket).
Flip the RX10K power switch to On while holding down the
ADJUST
adapter into appropriate AC power source (i.e., wall socket).
Flip the RX10K power switch to On while holding down the
ADJUST
▲ key.
2. Allow the RX10K to warm up for 15 minutes.
3. Connect a calibrated, high-accuracy voltmeter to the Voltage
Output terminals.
4. Scroll RX10K display to Input parameter screen. Press the
ADJUST
▲ key to scroll through the simulated inputs and
check the outputs.
Input
Output
+FS
10.000
± .010 VDC
Zero
0.000
± .005 VDC
-FS
-10.000
± .010 VDC
+FS/2
5.000
± .005 VDC
-FS/2
-5.000
± .005 VDC
+FS/4
2.500
± .005 VDC
-FS/4
-2.500
± .005 VDC
It is recommended that the system be checked for calibration
annually. If found to be out of specification, it can be returned to
Binsfeld Engineering Inc. for calibration for a nominal fee
($150.00, price subject to change).