E40.0.01.6C-02 Operating 
Manual 
EASY
LOG
 80 CL 
Page 6 of 12 
7.2  Operating possibilities at the device - logger operation 
Function 
Procedure 
Display and adjust 
cycle time 
Push button 
 and 
 at the same time. 
The main display shows “CYCL”. The secondary display 
shows the currently adjusted cycle time [in sec.]. 
If the logger is currently not recording, the desired cycle time 
[4 sec to 5h (18000 sec)] can be adjusted via the buttons 
 
 and 
 . 
Press button  
 again to confirm your adjustments and to 
leave the menu. 
Note:  
If no key is pressed for more than 2 min, the menu is 
quit without saving.  
Start recording 
Push button
 
 
for about 5 seconds.
 
The main display shows “run”.
  
The secondary display shows “no”.
 
To start recording select “YES” with the arrow keys 
(button 
 or 
 ) and verify with 
 . 
Note: 
If no key is pressed for more than 2 min, the menu is 
quit without saving.
 
Stop recording 
Push button 
 for about 5 seconds. 
The main display shows “HoLd”.  
The secondary display shows “no”
To stop recording select “YES” with the arrow key 
 (button 
 or 
 ) and verify with 
 . 
Note: 
If no key is pressed for more than 2 min, the menu is 
quit without saving. 
Notice: 
If you have activated one of the following options 
 • “logger stop via buttons disabled”  
 • “do not stop active logger”  
at the configuration of the logger the menu to 
stop the recording cannot be called. 
 
 
 
SET
1
2
2
3
SET
1
SET
1
2
3
SET
1
SET
1
2
3
SET
1