Greisinger EASYLOG 80 CL Humidity: 0 - 100 % r.H · Temperature: -25 to +60 °C · Pressure: 300 - 1100 hPa Temperature Dat 603414 Hoja De Datos
Los códigos de productos
603414
E40.0.01.6C-02 Operating
Manual
EASY
LOG
80 CL
Page 7 of 12
7.3 Operating possibilities via software GSOFT 40K
You can easily operate the logger via the software GSOFT 40K (> V7.14). This software
provides a variety of functions to operate the logger. For example:
• start and stop the logger
• read out the logger data; illustrate them graphically and save them
• adjust the alarm boundaries and alarm delay *
provides a variety of functions to operate the logger. For example:
• start and stop the logger
• read out the logger data; illustrate them graphically and save them
• adjust the alarm boundaries and alarm delay *
1
• activate the sea-level correction and enter the actual sea level *
1
Description:
The measured barometric air pressure can be converted to the air
pressure at sea level if the actual height above sea level is entered.
pressure at sea level if the actual height above sea level is entered.
• activate the function: “logger stop via buttons disabled“
*
*
1
Please note:
The alarm settings and the sea-level correction can be only
changed if the logger is stopped (= logger without logger data).
changed if the logger is stopped (= logger without logger data).
7.4 Operating possibilities via software EBxKonfig or EASYBus-
configurator
The software EBxKonfig (> V2.0) or EASYBus-configurator (> V3.9)
is needed to
change the configuration settings.
With that software the following adjustment are additionally possible.
• Selection of the unit of the calculated variables*
• Selection of the unit of the calculated variables*
2
• Offset and slope adjustment for each measured variable *
2
• Activation of the function: “do not stop active logger”
*
2
Please note:
This can be only changed if the logger is stopped (=logger without
logger data)
logger data)
8. Battery life time and recording period
Measuring cycle:
4 sec
3 min
15 min
5 h
Recording period:
11.5 days
521 days
7.1 years
142 years
Battery life time:
-- -- ca.
5
years
--
Please note:
Short measuring cycles seriously reduce the battery life time!
In this case you should consider not unplugging the EASYBus-
interface.
Then the logger is supplied via the interface.
The internal battery is spared.
In this case you should consider not unplugging the EASYBus-
interface.
Then the logger is supplied via the interface.