Control Functions
➭ Briefly press the ZERO | ESC key. 
The instrument acknowledges storage of the reference value 
with an acoustic signal, and the “ZERO 
ΔREL” or the “ΔREL” 
symbol appears at the LCD. The value measured at the 
moment the key is pressed serves as a reference value.
➭ The reference value can be cleared by once again pressing the 
ZERO | ESC key.
Notes Regarding Relative Measurement
• Relative measurement effects the digital display only. The 
analog display continues to read out the original measured value.
• In the case of relative measurement, 
Ω F or AC quantities may 
also appear as negative values.
4.3
Display (LCD)
4.3.1 Digital Display
Measured Value, Unit of Measure, Type of Current, Polarity
The measured value with decimal and plus or minus sign appears 
at the digital display. The selected unit of measure and current 
type are displayed as well. A minus sign appears to the left of the 
value during the measurement of zero-frequency quantities, if the 
plus pole of the measured quantity is applied to the “
⊥” input.
The “
0.d iSP” parameter can be used to determine whether lead-
ing zeros will be appear or be suppressed at the measured value 
display (see chapter 6.4).
Overranging
If the upper range limit of 3100 digits is exceeded “
0L” (overload) 
appears at the display.
Exceptions: “
0L” appears at the display as of 1000.0 V in the case 
of voltage measurement in the 1000 V range, as of 5.100 V for 
diode testing, and as of 11.00 A in the 10 A range.
4.3.2 Analog Display
Measured Value, Polarity
The analog display demonstrates the dynamic performance of a 
moving-coil mechanism. This display is especially advantageous 
for observing measured value fluctuation, and for balancing 
procedures.
Display mode Pointer: the current measured value is tracked in 
real-time.
The analog scale displays a negative range of 5 scale divisions for 
the measurement of zero-frequency quantities, allowing for 
precise observation of measured value fluctuation around zero. If 
the measured value exceeds the negative range of 5 scale 
divisions, polarity is reversed at the analog display. 
Scaling of the analog scale is automatic. This is very helpful for 
manual measuring range selection.
Overranging
Overranging in the positive range is displayed by means of the 
right triangle symbol.
Refresh Rate
In the bar graph and pointer modes, the analog display is 
refreshed 40 times per second.