Control Functions
4.4
Measured Value Storage: DATA (auto-hold / compare)
An individual measured value can be automatically “frozen” with 
the DATA function (auto-hold). This is useful, for example, when 
contacting the measuring points with the test probes requires 
your full attention. After the measuring signal has been applied 
and the measured value has settled in in accordance with the 
“condition” listed in the table below, the measured value is frozen 
at the digital display and an acoustic signal is generated. The test 
probes can now be removed from the measuring points, and the 
measured value can be read from the digital display. If the 
measuring signal falls below the value specified in the table, the 
function is reactivated for storage of the next value.
Measured Value Comparison (DATA Compare)
If the currently frozen value deviates from the first saved value by 
less than 100 digits, the acoustic signal is generated twice. If 
deviation is greater than 100 digits, only one brief acoustic signal 
is generated.
Note

The DATA function has no effect on the analog display, at 
which the current measured value continues to appear. 
However, when the digital display is “frozen”, the decimal 
point is fixed as well (fixed measuring range, symbol: MAN). 
The selected measuring range should not be manually 
changed as long as the DATA function is active.
The DATA function is deactivated by pressing and holding the 
DATA/MIN/MAX key (approx. 1 second), when the measuring 
function is changed or when the instrument is switched off and 
back on again.
1)
 Reactivation results from falling short of specified measured value limits.
2)
Two acoustic signals are generated the first time a measured value is saved as 
a reference value. For subsequent data hold, two acoustic signals are only 
generated if the currently frozen value deviates from the first saved value by less 
than 100 digits.
Key: MV = measured value, R = measuring range
DATA
Function
Press 
DATA /
Min/Max
Condition
Response from Instrument
Measuring 
Function
Measuring 
Signal
Display
Acou-
stic
MV
Digital
DATA
Activate
Brief
Blinks
Once
Save
(stabilized 
measured 
value)
V, A, F, Hz, 
%
> 10%  
of R
Is dis-
played
Static
Once
Twice 
2)
Ω 
 
≠ 0L
Reactivate 
1)
V, A, F, Hz, 
%
< 10%  
of R 
Stored 
MV
Blinks
Ω 
 
0L
Change to
Min/Max
brief
See table in chapter 4.4.1
Exit
long
Is cleared Is cleared
Twice
V, A, Hz
t [s]
100%
10%
Activated
Reactivated
Save
Save
3000 digits
of Measuring Range
F, %