hands-off measurements. 
  Do not use test leads or accessories that decrease voltage spacing. This diminishes arc 
protection and creates a hazardous condition. 
 
Use the following sequence when measuring high energy circuits:   
1.  De-energize the circuit using a  regular installed connect-disconnect device, such as a  circuit 
breaker, main switch, etc. 
2.  Attach the test leads to the circuit under test. Use appropriate safety rated test leads. 
3.  Set the multimeter to the proper measurement function and range. 
4.  Energize the circuit using the installed connect-disconnect device and make measurements 
without disconnecting the multimeter. 
5.  De-energize the circuit using the installed connect-disconnect device. 
6.  Disconnect the test leads from the circuit under test. 
 
WARNING:  The maximum common-mode voltage (voltage between COM  and the chassis 
ground) is 500V peak. Exceeding this value may cause a breakdown in insulation, 
creating a shock hazard. 
2.6.5    Power-on Defaults 
2831E and 5491B have a factory default setting for the power-on setting. 
 
Since the basic measurement procedures in this manual assume the factory defaults, reset the 
instrument to the factory settings when following step-by-step procedures. Table 2-2  lists the factory 
default settings. 
Table 2-2 Factory Default Settings 
 
Setting 
Factory Default 
Function 
Range 
Rate 
Remote/Local 
Trigger Mode 
Relative Mode 
Compare Mode 
HI Limit 
Lo Limit 
Percent Mode 
Reference 
Max/Min Mode 
Reading Hold 
Secondary Display Mode 
Cal Mode 
DCV 
AUTO 
Medium 
Local 
Immediate 
OFF 
OFF 
+1 
-1 
OFF 
+1 
OFF 
OFF 
OFF 
OFF