3.3    Measuring Current   
Model 2831E current measurement range: 2 mA, 20 mA, 200 mA, *2 A, *20 A 
Model 5491B current measurement range: 5 mA, 50 mA, 500 mA, *5 A, *20 A 
*Indicates range is available for manual range only.    Auto range will not work when reading requires 
setting to this range. 
 
Maximum resolution: 100nA (on 2 mA range (on 5 mA range for model 5491B)) 
Current measurement is available in Auto range for the lower current measurements only.     
 
Note: 
 
For 2831E, readings that are below 200 mA using the low current measurement input can be set in auto 
range.    This means it will auto range between 2 mA, 20 mA, and 200 mA range.    For readings above 
200 mA, auto range is not
 
 available.    Manual range must be used to read any readings higher than 200 
mA.    This means, for ranges 2 A and 20 A, only manual range is available.    Additionally, the high 
current measurement input must be used to obtain measurements in these two higher ranges. 
For 5491B, readings that are below 500 mA using the low current measurement input (fuse protected up 
to 500 mA max.) can be set in auto range.    This means it will auto range between 5 mA, 50 mA, and 
500 mA range.    For readings above 500 mA, auto range is not
 
 available.    Manual range must be used 
to read any readings higher than 500 mA.    This means, for ranges 5 A and 20 A, only manual range is 
available.    Additionally, the high current measurement input must be used to obtain measurements in 
these two higher ranges. 
3.3.1    Connections 
Assuming the multimeter is under factory default conditions, the basic procedure to measure current is 
as follows (users must use manual range for measurement, following step 3): 
1.  Connect test leads between 500 mA terminal (for lower current measurements) and COM terminal or 
between 20 A terminal (for higher current measurements) and COM terminal. 
2.  Select DCI or ACI measurement function by pressing 
  →   
 
or 
  →   
3.  Press 
  toggles auto ranging.  Note that auto range is only
4.  Connect test leads to the source as shown in Figure 3-2. 
  available for readings that are 
below 200 mA (for 2831E) and 500 mA (for 5491B) and when using the 500 mA max. terminal and 
COM terminal connections.    For readings above these limits, only manual range is available and 
the 20 A terminal and COM terminal must be used to obtain measurement results.    If instrument 
shows the  AUTO  annunciator  on display, press again to turn off auto ranging mode  and go into 
manual range mode.  Use the RANGE 
  and 
  keys to select a measurement range 
consistent with expected current. 
 
CAUTION:  Do not apply more than 1 A, 250 V to the 500 mA input terminal or the fuse will be 
blown.  Use 20 A terminal to measure current higher than 500 mA(or 200 mA for model 2831E).   
DCV 
ACV