Operating Manual GLF 100 RW 
Page 5 of 8 
 
 
8  Configuration of the instrument 
  To configure the instruments according to Your needs proceed like follows:: 
  Switch off instrument. 
  Switch on and press mode-key during the segment test (                  ), until the display shows 
the first parameter ‚P_oF„  
  If a parameter should be edited, press key up or down (              ), the setting of the parameter 
will be shown and can be changed via up/down keys and receipt the changed parameter with      
  Jump to the next parameter with key  
 
parameter 
value 
meaning 
key 
 
keys 
 
 
 
Auto Power-Off  
factory setting: Auto
 
 
Auto Power-Off in minutes. 
If there won‟t be pressed, the device will be 
switched off automatically to save battery power (adjustable range 1 to 120 
min) 
 
automatic power-off is deactivated (continuous operating) 
 
Unit of display  
factory setting: cond µS/cm
 
cond µS/cm 
Measuring of conductivity in unit µS/cm 
rESi MOhm*cm  Measuring of resistivity in unit MOhm*cm 
 
Range of display  
factory setting: Auto
 
Auto  
Measuring with auto-range 
100.0 µS/cm 
If Unit = cond µS/cm: conductivity range 0.0 ... 100.0 µS/cm  
20.00 µS/cm 
                 
“              : conductivity range 0.00 ... 20.00 µS/cm  
2.000 µS/cm 
                 
“              : conductivity range 0.000 ... 2.000 µS/cm (no auto-range) 
.2000 MOhm*cm  If Unit = rESi MOhm*cm: resistivity range 0.0100 ... 0.2000 MOhm*cm 
2.000 MOhm*cm                   
“                    : resistivity range 0.010 ... 2.000 MOhm*cm 
20.00 MOhm*cm                   
“                    : resistivity range 0.01 ... 20.00 MOhm*cm 
 
Unit of temperature display  
factory setting: °C
 
°C 
temperature is displayed in °C 
°F 
temperature is displayed in °F 
 
temperature compensation (only for µS/cm and mS/cm)
  factory setting: nLF
 
OFF 
No temperature compensation of conductivity measuring 
nLF 
non-linear temperature compensation for natural water acc. to EN 27888 
(DIN 38404) to measure ground water, surface water, drinking water and 
purest water 
Lin 
Linear temperature compensation with adjustable coefficient (t.Lin) 
nA.CL 
Temperature compensation of weak NaCl-solutions acc. to DIN EN 60746-3 
 
Reference temperature of temp. compensation (only for t.Cor <> oFF)  
factory setting: 25 °C
 
25 °C / 77 °F 
Reference temperature 25 °C / 77 °F 
20 °C / 68 °F 
Reference temperature 20 °C / 68 °F 
 
Compensation coefficient (only for t.Cor = Lin)  factory setting: 1.000 
0.300 .. 5.000 
Temperature compensations coefficient in % / K. 
The factor has to be evaluated by experiment or resolved from literature. 
 
Restore to factory settings 
 
Settings are kept 
 
Restoring device to factory settings 
 
Pressing         again stores the settings, the instruments restarts (segment test) 
Please note:  If there is no key pressed within the menu mode within 2 minutes, the configuration 
will be cancelled, the entered settings are lost!