Bacharach Ammonia Gas Monitor Single-Zone Manual De Propietario

Descargar
Página de 40
Single Zone Gas Monitor 
P/N: 3015-4256 Rev 13 
19 
3.3.  Display Screens 
3.3.1.  Initial Power Up 
When the SZ monitor is first powered up all front panel lights are turned 
ON, and a splash screen appears showing the monitor’s firmware 
revision level. After a brief moment the Warm Up screen is display 
along with the front panel MONITOR ON light (green) blinking. 
The monitor takes 15 minutes to warm up; after which, the MONITOR 
ON
 light glows steadily and the Data Display screen is displayed. 
 
 
3.3.2.  Data Display Screen 
During normal operation, the Data Display  screen shows when the 
monitor is performing the following three functions: 
MEASURE indicates that the monitor is actively measuring for selected 
gas. In the example screens shown above, the monitor is currently 
detecting 0 ppm of R-134A (refrigerant gas), and that a peak 
measurement of 485 ppm has occurred.  
To reset the peak value to zero, press the ENTER button. 
PURGE  is displayed when the monitor is resetting its infrared detector 
to a baseline of 0 ppm using fresh (zero) air that is being drawn in 
through the Purge Air Inlet Port. This purging process is performed on 
an “as needed” basis, which is normally once every 6 to 8 minutes. 
 
 
 
PRES CHK is displayed when the monitor is performing an atmospheric pressure check, which is done to 
ensure the accuracy of the gas measurement under varying atmospheric conditions. This pressure check 
is performed every 30 minutes. 
3.3.3.  Function Screens 
The Function screens are used to display stored data and to set up the monitor. 
From the Data Display  screen, press any Keypad  button to display the first Function Menu  screen. 
Next, use the Keypad buttons to move the arrow (>) on the display until it is next to the desired function, 
and then press the ENTER button to select that function.  
   
   
 
   
 
 
Once a function has been selected, use the Keypad to scroll through the displayed data or to change a 
parameter associated with that function. Press ENTER  to save newly entered parameters. Press the 
QUIT button to return to the previous screen without saving. 
Note that if no buttons are pressed within 90 seconds after selecting a function, the monitor returns to the 
Data Display screen. 
 
>SILENCE ZONETMP 
 SQUELCH CAL 
 
>P-CHK 
  
 
>LOG INT AUDALRM 
 LOOP    CLOCK 
 
PRES CHK 00485pk 
    0ppm R134A 
 
PURGE    00485pk 
    0ppm R134A 
MEASURE  00485pk 
    0ppm R134A 
WARM UP 
    BACHARACH   
  VERSION x.xx 
>PPM LOG FAULTS 
 ALARMS  DIAGNOS 
>GASTYPE SPILLVL 
 LEAKLVL EVACLVL