SECTION 4.  MAINTENANCE 
4.1.  Warnings and Cautions 
 
WARNING:  Shock Hazard!  Always disconnect AC power before working inside the 
monitor. 
 
 
WARNING:  Before performing any operation or maintenance, verify connection of the AC 
power ground wire to the monitor’s ground stud.  Under no circumstances should this 
monitor be operated  or serviced  without a protective ground.  Doing so poses a potential 
shock hazard and is also a violation of electrical safety standards applicable to this type of 
equipment. 
 
 
WARNING:  Electrical installation should be performed by a certified electrician and must 
comply with all applicable NEC/CEC and local electrical safety codes. 
 
 
CAUTION:  When working inside the monitor, be very careful not to dislodge any electrical 
wiring or pneumatic tubing. The SZ monitor contains sensitive electronic components that 
can be easily damaged. Be careful not to touch or disturb any of these components. 
 
4.2.  Charcoal Filter (Halogen Gas Monitor Only) 
During normal operation, the Single Zone monitor performs a purge cycle during 
which fresh air is drawn in from the purge line to flush internal components of any 
residual refrigerants. Purge lines should draw in fresh outside  air to perform the 
purge  function.  When access to outdoor air is impractical, purge lines may begin 
indoors, however a charcoal filter must be used on the purge line to ensure that no 
contaminated air is drawn into the system during purge cycles.  The charcoal filter 
(which is mounted outside of  the enclosure using the included bracket)  removes 
refrigerant gas from the purge-air stream during the internal purging process. 
 
External Filter Includes 
Mounting Bracket 
Use upgrade kit 3015-5716 to convert an SZ with internally mounted filter to an SZ with an external filter.
 
Replace the charcoal filter (P/N 3015-3125) about every 6 months, when a zero filter fault occurs (fault 
code <0100>), or after the monitor itself has been exposed to unusually high levels of refrigerant gas, 
such as after an evacuation alarm. 
4.3.  Hydrophobic Filters 
Hydrophobic  Filters  (see  Figure  2-5  on page 11  and  Figure  2-6  on page 12)  located  in both the gas-
sample and purge lines prevent water from entering the IR detector. A zone flow fault will occur (fault 
code <0800>) if the gas-sample filter becomes blocked, while a purge flow fault will occur (fault  code 
<1000>) if the purge filter becomes blocked. Replace the appropriate filter as required (P/N 0007-1650). 
4.4.  Servicing Air Lines and Termination Filters 
The gas-sample line and the optional purge and exhaust lines should be periodically checked for obvious 
signs of kinks, damage, and contamination. Replace the tubing as required. 
The gas-sample line and purge-line  termination filters  prevent dust and dirt from entering the monitor. 
Both  of these filters should be periodically checked and replaced when there are obvious signs of 
contamination. A zone flow fault will occur (fault code <0800>) if the gas-sample line filter becomes 
blocked, while a purge flow fault will occur (fault code <1000>) if the purge-line filter becomes blocked. To 
service the filter, simply remove it from the line and replace it with a new one (P/N 3015-3420).