Frye WinCHAP ver. 2.82 Manual De Usuario

Descargar
Página de 124
WinCHAP User’s Manual 
Test Menu   
45
4.2  WinCHAP Measurements
Each measurement window in WinCHAP is either attached to the hearing aid, 
or it is attached to the client:
•  Client: Audiometric Test, DSL/NAL Test.
•  Hearing Aid: ANSI 96, ANSI 87, ANSI 92, ACIC, IEC, JIS, Response 
Curves, and Real Ear.
4.2.1  Creating a client test
1.  Click Client-Patient in the left column of the Test Menu to enter the 
Client-Patient window.
2.  Locate the record of the client that you want to test. Use the Search but-
ton if necessary. (See Section 4.1.2.3.)
3.  Click the name of the client in the right column of the Test Menu. This 
will bring up a window containing the client test options.
4.  Double-click the test you would like to create. See Chapter 5 for details 
on using the Audiometric Test window, and Chapter 7 for details on using 
the DSL/NAL Test window.
4.2.2  Creating a hearing aid test
1.  Locate the record of the hearing aid that you want to test. If the hearing 
aid is attached to a client, it might be more convenient to find the client 
record and click the PLUS (+) next to his name. This will display all tests 
and hearing aids attached to that client.
2.  Double-click the hearing aid record in the right column of the Test Menu. 
This will bring up a window containing all the hearing aid test options.
3.  Double-click the test you would like to create. See Chapter 6 for details 
on using the hearing aid testing windows.
4.2.3  Opening an existing test
1.  Locate the hearing aid or client record to which the test is attached.
2.  Click the PLUS (+) next to the client or hearing aid record. This will 
list all attached tests. (If you are viewing a client record, you may see an 
additional PLUS (+) next to any hearing aid attached to that client. Click 
it to view the tests saved to that hearing aid.)
3.  Double-click the existing test to open it.
4.2.4  Deleting a test
You can delete any measurement attached to a hearing aid or client. Entire hear-
ing aid or client records must be deleted in the Office Manager.
1.  Select the test you would like to delete.
2.  Click the 
Delete button in the Test Menu toolbar.