2.1.1 Sensor Installation or Replacement 
Note:  The R17MED oxygen sensor must be installed before the 
oxygen analyzer/monitor can be operated  
1.  Remove the new sensor from its protective bag. Inspect the 
sensor for damage or electrolyte leakage. If the sensor is 
damaged, obtain a replacement. Do not use the defective 
sensor as it may damage the unit. 
WARNING: 
THE SENSOR ELECTROLYTE IS CAUSTIC. DO NOT 
LET IT COME IN CONTACT WITH SKIN.  IF IT DOES, 
FLUSH AFFECTED AREA WITH WATER. DO NOT 
ATTEMPT TO OPEN OR REPAIR THE SENSOR.  
WARNING: 
THE SENSOR ALSO CONTAINS LEAD. LEAKING OR 
EXHAUSTED SENSORS SHOULD BE HANDLED AND 
DISPOSED OF IN ACCORDANCE WITH LOCAL 
REGULATIONS. CONTACT TELEDYNE FOR THE 
MATERIAL SAFETY DATA SHEET  
2.  Plug one end of the coiled cable into the jack receptacle on the 
back end of the R17MED sensor and secure in place with the 
capture nut located at the base of the connector. See Figure 2-1. 
 
Figure 2-1: Installing the R17MED Sensor 
 
3.  Plug the other end of the coiled cable into the receptacle located on 
the right side of the unit and secure it in place using the capture nut. 
See Figure 2-2.