Proline Promass 80 PROFIBUS PA
Commissioning
47
6.3
Quick Setup
In the case of measuring devices without a local display, the individual parameters and functions 
must be configured via the operating program, e.g. FieldCare.
If the measuring device is equipped with a local display, all the important device parameters for 
standard operation, as well as additional functions, can be configured quickly and easily by means 
of the following Quick Setup menus.
6.3.1
Quick Setup "Commissioning"
a0004749-en
Fig. 31: 
Quick Setup "Commissioning"
+
+
+
E
E
HOME-POSITION
Esc
E
+
-
XXX.XXX.XX
Yes
Yes
No
No
Mass flow
Volume flow
Density
Selection System Units
Configuration another unit ?
Automatically configuration display ?
Temperature
Quit
Unit
Mass flow
Unit
flow
Volume
Unit
Density
Unit
Temperature
Unit
Totalizer
Unit
Totalizer
Automatically
parameterization
of the display
Language
Pre-setting
Quick Setup
Quick Setup
Commission
n
o
Corr. Vol. flow
Unit
Corr. Vol. flow
Unit
Ref. Density
Fix
Ref. Density
p
Selection pre-settings
Actual Settings
Deliver Settings
y
m
Set Unit to Bus
SET UNITS
OFF
m
!
Note! 
• The display returns to the cell SETUP COMMISSIONING (1002) if you the press 
X key combination during 
parameter interrogation. The stored parameters remain valid.
• The "Commissioning" Quick Setup must be carried out before one of the Quick Setups explained below is run.
m The "DELIVERY SETTINGS" option sets every selected unit to the factory setting.
The "ACTUAL SETTINGS" accepts the units you configured beforehand.
n Only units not yet configured in the current Setup are offered for selection in each cycle. The unit for mass, volume 
and corrected volume is derived from the corresponding flow unit.
o The "YES" option remains visible until all the units have been configured. "NO" is the only option displayed when 
no further units are available.
p The “automatic parameterization of the display” option contains the following basic settings/factory settings:
YES: line 1= mass flow; line 2 = totalizer 1
   NO: The existing (selected) settings remain.