Agilent Technologies 8935 series e6380a Manuale Utente

Pagina di 320
Chapter 4
113
Overview of the Test Set’s Built-In Tools
Using the Spectrum Analyzer
Using the Spectrum Analyzer’s Delta Markers
Delta markers are displayed when the 
Norm/Delta
 field on the 
spectrum analyzers 
Marker
 menu is set to 
Delta
.
The delta makers are positioned in the same way as the normal 
markers using the 
Marker
 menu of the spectrum analyzer screen. See 
When the delta maker is activated, the 
Marker To Peak
Next Peak
Center Freq
, and 
Ref Level
 fields control only the delta marker. 
Peak
 or 
Next Peak
 - The normal marker remains stationary. When a 
delta marker to center frequency measurement would move the 
reference marker off the screen, the measurement is aborted and the 
following message is displayed: 
Attempt to move Reference 
Marker off the screen
.
Center Freq
 or 
Ref Level
 - The normal marker moves as necessary 
to stay at the same signal frequency when the spectrum display 
changes.
Delta Frequency
The delta marker frequency measurement (
Delta Mrkr Freq
) is the 
frequency of the delta marker minus the frequency of the normal 
marker. If the delta marker is to the right of the normal marker, the 
delta frequency is positive. If the delta marker is to the left of the 
normal marker, the delta frequency is negative. 
The units for this measurement are GHz, MHz, kHz, and Hz.
Delta Level
The delta marker level measurement (
Delta Mrkr Lvl
) is the level of 
the normal marker minus the level of the delta marker in dBm. If the 
delta marker is higher than normal marker, the delta level is positive. 
If the delta marker is lower than the normal marker, the delta level is 
negative. 
The units for this measurement are dB. (There is a percent unit 
available; however, since the spectrum analyzer level is always 
displayed on a logarithmic scale, linear units are inappropriate.)