Chapter 5
153
CDMA Measurements - Screens and Control Fields
Control Fields for CDMA Measurements
Anl Special
Analyzer special modes change how IQ modulated signals are analyzed.
Normal
 sets the CDMA generator and CDMA analyzer to create and 
analyze signals with standard rotation. 
Inverted
 sets reverse rotation for I and Q signals; opposite to the 
direction specified in the IS-95 standards. In this mode, the CDMA 
generator creates reverse-rotation IQ signals and the CDMA 
analyzer expects reverse-rotation IQ signals.
Screens on which this field is present
CDMA ANALYZER
Auto Zero
This field is displayed for average power measurements.
Auto
 is used to automatically zero the average power measurement 
periodically during operation.
Manual
 is used to override the 
Auto
 feature. When 
Manual
 is selected, 
you must manually zero the average power measurement by selecting 
the 
Pwr Zero
 field.
Screens on which this field is present
CDMA ANALYZER
Carrier FT or Carrier Feedthru
This field is a measurement display field. See 
 for more information about this measurement.
Screens on which this field is present
COD DOM (Code Domain Analyzer)
CDMA ANALYZER
CDMA GENERATOR
Chan Num
IS-2000 only. Indicates the supplemental channel and Walsh channel or 
code (in parenthesis), at which the marker is currently located. The 
field will change when you change the value of the marker position 
using the numeric keypad or the knob in Marker control mode with 
Marker Pos is selected.
Screens on which this field is present
COD DOM (Code Domain Analyzer, IS-2000 only)