Agilent Technologies 8935 series e6380a Manuale Utente
158
Chapter 5
C:\Spk\Ref\RefGuideRevE\CdmascrnE2.fm
CDMA Measurements - Screens and Control Fields
Control Fields for CDMA Measurements
Even Sec In
This field controls the internal connection of the EVEN SECOND
SYNC IN connector to the CDMA analyzer or generator.
SYNC IN connector to the CDMA analyzer or generator.
• When
Enable
is selected, signals applied to the EVEN SECOND
SYNC IN connector synchronize the timing circuits.
• When Not is selected, input signals applied to the EVEN SECOND
SYNC IN are ignored. This mode may be useful when signals other
than an even-second clock are used for synchronization. For
example, when using a one-second clock, you would select
than an even-second clock are used for synchronization. For
example, when using a one-second clock, you would select
Not
after
synchronization to prevent the internal even-second clock from being
retriggered half-way through its cycle.
retriggered half-way through its cycle.
Screens on which this field is present
CDMA ANALYZER
CDMA GENERATOR
CDMA GENERATOR
Find PN
This field selects whether the PN offset will be automatically calculated
(
(
Auto
), or manually entered (
Manual
). (See
and
for more information.)
Screens on which this field is present
CODE DOM (Code Domain Analyzer)
Freq Err
This measurement display field displays the frequency error. See
for more information about
this measurement.
Screens on which this field is present
CODE DOM (Code Domain Analyzer)
CDMA ANALYZER
CDMA GENERATOR
CDMA ANALYZER
CDMA GENERATOR