CDMA Measurements - Screens and Control Fields
Control Fields for CDMA Measurements
Gain
This field sets the IF gain of the CDMA analyzer for rho, EVM, and 
channel power measurements. (Range: 0 dB to 36 dB in 6 dB steps.)
Gain is adjusted two ways: 
Auto
 measures the signal level and automatically adjusts the gain. 
Hold
 disables automatic operation to allow you to manually enter 
the desired gain by selecting the current gain setting and entering a 
new value using the keypad or cursor control knob.
This field is replaced by the 
Pwr Gain
 field when measuring average 
power on the CDMA ANALYZER screen. See 
 
for more information.
Screens on which this field is present
CDMA ANALYZER
COD DOM (Code Domain Analyzer): gain controls menu
Gen Dir
This field sets the link direction of the CDMA signal. 
Fwd
 (forward) sets the CDMA generator to produce QPSK 
modulation to imitate a base station. The EQ filter field is 
automatically displayed and set to 
EQ In
 to apply an equalizing 
filter to the I and Q paths as part of the baseband filtering, but it can 
be changed to 
Out
Rev
 (reverse) sets the CDMA generator to produce OQPSK 
modulation to imitate a mobile station. The EQ filter field is not 
displayed because it is automatically set to 
Out
 (and cannot be 
changed). It is set to 
Out
 to remove the equalizing filter from the I 
and Q paths. 
Screens on which this field is present
CDMA GENERATOR