Chapter 5
161
CDMA Measurements - Screens and Control Fields
Control Fields for CDMA Measurements
Input Atten
This field controls the RF attenuators of the Test Set. If an 
ADC 
Overdriven
 message is displayed at the top of the screen while making 
measurements, increase the attenuator setting until the message is no 
longer displayed. This is the coarse level adjustment for the signal level 
into the CDMA analyzer. Fine level adjustment is performed using the 
Gain
 field. 
See 
 for more information.
NOTE
The 
Input Atten
 setting does not affect, and is not displayed with, the 
average power measurement.
Screens on which this field is present
CDMA ANALYZER
CODE DOM (Code Domain Analyzer): gain controls menu
Input Port 
This field selects which Test Set port to use as the RF input. This field 
is not displayed when measuring average power, since you must use the 
RF IN/OUT port to measure average power. 
CAUTION
Instrument Damage 
To prevent instrument damage, do not exceed the maximum power level 
labeled at the connectors on your Test Set. The ANT IN connector is for 
measuring low level RF signals. High level (>60 mW) transmitter 
signals must be measured using the RF IN/OUT port. 
Screens on which this field is present
CDMA ANALYZER
CODE DOM (Code Domain Analyzer): main controls menu
Lvl
This measurement displays the relative level of the selected Walsh 
channel when compared to the total power in a 1.23 MHz bandwidth 
centered around the 
Tune Freq
 setting. The measurement is displayed 
when the 
code domain analyzer’s main
 controls 
Measurement
 field 
is set to 
Power
 or 
Fast Pwr
. This function is appears as 
Marker Lvl
 for 
the IS-2000 option. See 
Screens on which this field is present
COD DOM (Code Domain Analyzer, IS-95 only): Marker controls menu