Chapter 5
165
CDMA Measurements - Screens and Control Fields
Control Fields for CDMA Measurements
Phse
This field displays the results of the code domain phase measurement. 
This field is displayed on the CODE DOM screen when the 
Measurement
 field is set to 
Phase
For more information about the code domain phase measurement see 
.
Screens on which this field is present
CODE DOM (Code Domain Analyzer, IS-95 only)
PN Incrment
This value is used to calculate the PN offset for the system. This field is 
displayed for EVM or Rho measurements when the 
Find PN
 field is set 
to 
Auto
If you know the PN increment for the system you are testing, you can 
enter it in this field (this speeds the PN offset calculation). Enter 
1
 in 
this field if the PN increment is unknown.
Screens on which this field is present
CDMA ANALYZER
CODE DOM (Code Domain Analyzer): PN setup controls menu
PN Offset
This field is displayed for EVM or Rho measurements when the 
Find PN
 field is set to 
Manual
Use this field to enter the PN sequence offset index for your base 
station. Each whole offset is equal to 64 chips (= 52.08 
µ
s). Fractional 
values are rounded off, and can be entered, in increments of 0.015625 
(1 chip). Once entered, this value is shared by all CDMA screens.
This value is used by the analyzer when the 
Even Sec In
 field on the 
CDMA GENERATOR screen is set to 
Enable
If the PN offset is unknown, the Test Set can calculate the PN offset 
using the PN increment. See 
NOTE
When the 
Even Sec In
 field is set to 
Not
, changes to the 
PN Offset
 
field have no effect on the analyzer.
Screens on which this field is present
CDMA ANALYZER
CODE DOM (Code Domain Analyzer): PN setup controls menu