Agilent Technologies 8935 series e6380a Manuale Utente
166
Chapter 5
C:\Spk\Ref\RefGuideRevE\CdmascrnE2.fm
CDMA Measurements - Screens and Control Fields
Control Fields for CDMA Measurements
PN Offset or PN Ofs
This measurement display field (PN Offset) is displayed on the CDMA
ANALYZER and CDMA GENERATOR screens when Rho
measurements are selected. If the PN Offset measurement is not
displayed, select the Freq Err measurement field, the choose PN Offset.
The same measurement is displayed in the
ANALYZER and CDMA GENERATOR screens when Rho
measurements are selected. If the PN Offset measurement is not
displayed, select the Freq Err measurement field, the choose PN Offset.
The same measurement is displayed in the
PN Ofs
measurement field
on the CODE DOM screen.
See
for more information.
NOTE
When the
Even Sec In
field is set to
Not
, changes to the
PN Offset
field have no effect on the analyzer.
Screens on which this field is present
CDMA ANALYZER
CDMA GENERATOR
CODE DOM (Code Domain Analyzer): PN setup controls menu
CDMA GENERATOR
CODE DOM (Code Domain Analyzer): PN setup controls menu
Pwr Gain
This field is used to set the level into the CDMA analyzer when making
average power measurements. Levels are set in 6 dB increments. The
IF signal level is displayed as
average power measurements. Levels are set in 6 dB increments. The
IF signal level is displayed as
ADC FS
when average power is measured,
and should be between -1.0 and- 5 dB for best measurement results.
•
Auto
automatically sets the gain for an ADC FS value in the range of
−
1 to
−
5 dB. This is the recommended mode of operation.
•
Hold
lets you override the automatic setting to manually set the
gain.
This field is only displayed when an average power measurement is
selected.
selected.
Screens on which this field is present
CDMA ANALYZER