Agilent Technologies 8935 series e6380a Manuale Utente

Pagina di 320
166
Chapter 5
C:\Spk\Ref\RefGuideRevE\CdmascrnE2.fm
CDMA Measurements - Screens and Control Fields
Control Fields for CDMA Measurements
PN Offset or PN Ofs
This measurement display field (PN Offset) is displayed on the CDMA 
ANALYZER and CDMA GENERATOR screens when Rho 
measurements are selected. If the PN Offset measurement is not 
displayed, select the Freq Err measurement field, the choose PN Offset. 
The same measurement is displayed in the 
PN Ofs
 measurement field 
on the CODE DOM screen.
 for more information.
NOTE
When the 
Even Sec In
 field is set to 
Not
, changes to the 
PN Offset
 
field have no effect on the analyzer.
Screens on which this field is present
CDMA ANALYZER
CDMA GENERATOR
CODE DOM (Code Domain Analyzer): PN setup controls menu
Pwr Gain
This field is used to set the level into the CDMA analyzer when making 
average power measurements. Levels are set in 6 dB increments. The 
IF signal level is displayed as 
ADC FS
 when average power is measured, 
and should be between -1.0 and- 5 dB for best measurement results. 
Auto
 automatically sets the gain for an ADC FS value in the range of 
1 to 
5 dB. This is the recommended mode of operation.
Hold
 lets you override the automatic setting to manually set the 
gain.
This field is only displayed when an average power measurement is 
selected. 
Screens on which this field is present
CDMA ANALYZER