Agilent Technologies 8935 series e6380a Manuale Utente

Pagina di 320
Chapter 5
169
CDMA Measurements - Screens and Control Fields
Control Fields for CDMA Measurements
field on the CODE DOMAIN ANALYZER screen. 
Ampl Mid
 qualifies a trigger each time the input to the CDMA 
analyzer rises above 
12 ADC FS. 
Ampl Hi
 qualifies a trigger each time the input to the CDMA 
analyzer rises above 
− 
6 ADC FS. 
External
 must be selected if the trigger qualifier is provided by an 
external source. The external signal connects to the TRIGGER 
QUALIFIER IN connector. A qualifier occurs on the input signal’s 
rising edge. 
Screens on which this field is present
CDMA ANALYZER
CODE DOM (Code Domain Analyzer): trigger controls menu
Rate RC-3
This field appears only in IS-2000 Code Domain measurements. The 
field displays the data rate of the currently selected channel in terms of 
an RC3 radio configuration. 
Screens on which this field is present
CODE DOM (Code Domain Analyzer, IS-2000 only)
Rate RC-4
This field appears only in IS-2000 Code Domain measurements. The 
field displays the data rate of the currently selected channel in terms of 
an RC4 radio configuration. 
Screens on which this field is present
CODE DOM (Code Domain Analyzer, IS-2000 only)
Rate RC-5
This field appears only in IS-2000 Code Domain measurements. The 
field displays the data rate of the currently selected channel in terms of 
an RC5 radio configuration. 
Screens on which this field is present
CODE DOM (Code Domain Analyzer, IS-2000 only)