Agilent Technologies 8935 series e6380a Manuale Utente

Pagina di 320
168
Chapter 5
C:\Spk\Ref\RefGuideRevE\CdmascrnE2.fm
CDMA Measurements - Screens and Control Fields
Control Fields for CDMA Measurements
Pwr Zero
Selecting this field calibrates (zeroes) the average power meter. This 
operation should be performed immediately before making an average 
power measurement. This field is only displayed if the average power 
measurement is selected. 
Operating Considerations
The 
Auto Zero
 field causes average power to be periodically calibrated 
when set to 
Auto
. If the 
Auto Zero
 field is set to 
Manual
, you must use 
the 
Pwr Zero
 field to zero average power. (See 
for more information.)
Screens on which this field is present
CDMA ANALYZER
Q Power Level
This IS-2000 Code Domain display field shows the power (in the lower 
bar graph) of the currently selected Q (IS-2000) Walsh code. 
Screens on which this field is present
COD DOM (Code Domain Analyzer, IS-2000 only): Complex Power 
measurement.
Qual Event
This field selects a trigger qualifier. The trigger qualifier serves as a 
precursor to the analyzer’s trigger event. Selecting a trigger qualifier is 
required when a time-delayed trigger event is desired, or if the trigger 
signal is externally supplied at the TRIGGER QUALIFIER IN 
connector. 
If a time-delayed trigger is used (see 
), the 
qualifier starts the delay timer. When the timer expires, a trigger event 
occurs and a measurement is made. 
None
 causes the trigger event to be determined solely by the 
selection in the 
Trig Event
 field. 
27 ms
 is an internally generated clock signal. 
20 ms
 is an internally generated clock signal. 
80 ms
 is an internally generated clock signal (default selection). 
2 s
 is an internally generated clock signal.
Ampl Lo
 qualifies a trigger each time the input to the CDMA 
analyzer rises above 
18 dB ADC FS. 
ADC FS
 is displayed when the 
average power measurement is selected, or when adjusting the 
Gain