Chapter 6
213
Analog Measurements - Screens and Control Fields
Control Fields for Analog Measurements
RF Gen Freq
This field sets the RF generator’s frequency. 
This field is displayed on the screens listed below when the 
RF Display
 
field on the INSTRUMENT CONFIGURE screen is set to 
Freq
.
Screens on which this field is displayed
SPEC ANL: RF Gen controls menu
RF GENERATOR
RF In/Ant
This field selects the input port for the spectrum analyzer. 
Operating Considerations
Maximum signal levels are printed on the connector panel. 
CAUTION
Connecting a signal of >60 mW to the ANT IN port can cause 
instrument damage (although internal protection circuits can typically 
withstand a short-duration signal of several Watts). 
If the overpower circuit is triggered (signified by a warning message at 
the top of the screen), remove the signal from the ANT IN port, and 
reset using 
Meas Reset
 (press and release the 
Shift
 key, then the 
Hold
 
key) or turn the Test Set off and on to reset it.
Using the ANT IN port with the 
Sensitivity
 field set to 
High
 can 
result in uncalibrated operation (a message appears on the screen when 
this happens). The purpose for the high sensitivity setting is to allow 
you to look and listen to very low level signals when absolute accuracy 
is not essential. 
See 
 for more 
information.
Screens on which this field is displayed
SPEC ANL