Chapter 7
C:\Spk\Ref\RefGuideRevE\Cnfgscrn.fm
Configuration - Screens and Control Fields
Control Fields on the Configuration Screens
RF In/Out
This field is used to indicate losses or gains between the RF IN/OUT 
port and the device-under-test. 
• Enter  a  positive value to indicate a gain (such as an amplifier gain). 
When the RF IN/OUT port is used as an output, the RF generator’s 
(or tracking generator’s) level is automatically set the specified 
amount below what is indicated in the RF generator’s 
Amplitude
 
field. Example; if this value is 10 dB, and the 
Amplitude
 field shows 
0 dBm, the actual level out of this port is 
−1
0 dBm.
When this port is used as an input, the 
TX Power
 measurement and 
spectrum analyzer’s marker level (
Lvl
) are automatically reduced by 
that amount. 
• Enter  a  negative value to indicate a loss (such as cable loss). The RF 
generator’s (or tracking generator’s) level out this port is 
automatically set that amount above what is indicated in the RF 
generator’s 
Amplitude
 field. 
When used as an input, the 
TX Power
 and the spectrum analyzer’s 
marker level (
Lvl
) measurements are increased by that amount. 
This field is only used when the 
RF Level Offset
 field is set to 
On
 for more information.
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE
RF Level Offset
This field enables/disables the RF level offsets entered in the 
RF 
In/Out
Duplex Out
, and 
Antenna In
 fields below it. 
• When set to 
On
, the RF generator’s amplitude and RF analyzer’s 
power measurement are offset by the values entered in these fields. 
• When set to 
Off
, the values in these fields are ignored. 
For more information see the following related topics:
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE