Chapter 7
247
Configuration - Screens and Control Fields
Control Fields on the Configuration Screens
RF Offset
This field is displayed when the 
RF Display
 field is set to 
Freq
This field enables/disables the RF generator 
− 
RF analyzer frequency 
offset specified in the 
(Gen)-(Anl)
 field below it. 
Operating Considerations
When an RF offset is used, changing the RF generator’s frequency or 
RF analyzer’s tune frequency automatically alters the other setting. 
See 
 for more 
information.
Screen(s) Where Field is Present
INSTRUMENT CONFIGURE
Save/Recall 
This field specifies which memory device the Test Set accesses when the 
save and recall functions are used. 
Internal
 is a section of internal RAM. RAM is also used for running 
IBASIC programs, which may require you to delete the save/recall 
registers if the program is very large. 
Card
 is the front-panel MEMORY CARD (PC card) slot. A 
Save/Recall Device is not Present 
message is displayed if you 
try to save or recall an instrument setup when a write-able PC card 
is not installed in the Test Set. 
RAM
 refers to RAM disks that you can create on internal RAM. Refer 
to the Programmer’s Guide for information on creating RAM disks. 
This is part of the same memory used when “internal” is specified, 
and may have to be erased when loading very large IBASIC 
programs. A 
Save/Recall Device is not initialized
 message is 
displayed if you try to save or recall an instrument setup when a 
RAM disk has not been created. 
Screen(s) Where Field is Present
I/O CONFIGURE