Agilent Technologies E6380-90073 Manuale Utente
![Agilent Technologies](https://files.manualsbrain.com/attachments/f20ae6efbee7636e6d0d1a481fd9e3847e3caeb5/common/fit/150/50/991587f279eee8cef3aa0a068e5c2f53d86b00e1bf17fbcbf9a87aa49d96/brand_logo.png)
102
Chapter 1
GPIB Command Dictionary
CANalyzer subsystem
:CHANnel:POWer:FILTer
:CHANnel:POWer:FILTer?
:CHANnel:POWer:FILTer?
These commands set/query the filter for the channel power
measurement.
measurement.
• ‘1.23M’ measures the absolute channel power level of the RF signal
in a 1.23 MHz bandwidth centered around the RF channel or tune
frequency setting. This measurement must be calibrated using the
CAN:CHAN:CPOW:CAL command each time the frequency is
changed.
frequency setting. This measurement must be calibrated using the
CAN:CHAN:CPOW:CAL command each time the frequency is
changed.
• ‘30kHz’ measures the adjacent channel power (that is the power in
the sidebands around the chosen CDMA channel).
Syntax
CAN:CHAN:POW:FILT ‘1.23M’
‘30kHz’
CAN:CHAN:POW:FILT?
Screen/field equivalent
CAN:CHAN:POW:FILT controls the
Ch Pwr Filt
field of the CDMA
ANALYZER screen while measuring channel power.
:DARM
This command works with the :ARM command to control the
measurements.
measurements.
Syntax
CAN:DARM
Screen/field equivalent
CAN:DARM controls the
Disarm
subfield of the
Analyzer
field on the
CDMA ANALYZER screen.