Chapter 1
123
GPIB Command Dictionary
CDANalyzer subsystem
:GAIN:MODE <string>
:GAIN:MODE?
These commands set/query the mode of the 
Gain
 field, allowing you to 
choose between an autoranging gain (‘Auto’) or a fixed gain (‘Hold’).
This command may conflict with the CDMA analyzer’s settings 
CAN:PATH:GAIN and CAN:PATH:MODE. For example, if autoranging 
is turned off (‘Hold’) in the CDMA analyzer and turned on in the Code 
Domain Analyzer (‘Auto’), the ‘Hold’ setting will always override the 
‘Auto’ setting. When autoranging is turned off, setting either analyzer’s 
path gain will automatically change the other’s to the same setting.
For average power measurements, use the CAN:POW:GAIN:MODE 
command.
Syntax
CDAN:GAIN:MODE ‘Auto’
   
‘Hold’
CDAN:GAIN:MODE?
Screen/field equivalent
CDAN:GAIN:MODE sets the 
Auto/Hold
 subfield of the 
Gain
 field of the 
CODE DOM screen, when the CDMA standard is set to “IS-95 Only”.
:GAIN:VALue?
This command queries the gain of the code domain analyzer. This 
command will only be enabled when autoranging is turned on 
(CDAN:GAIN:MODE ‘Auto’). 
Syntax
CDAN:GAIN:VAL? !returns 0 dB,6 dB,12 dB,18 dB,24 dB, 30 db, 
36 dB
Screen/field equivalent
When the CDMA standard is set to “IS-95 Only”, the 
CDAN:GAIN:VAL? reads the lower subfield of the 
Gain
 field of the 
CODE DOM screen when the upper subfield is set to 
Auto